TSH FIA Test
Quantitative, non-subjective results.
Increased sensitivity and broader detection range compared to rapid tests.
RANGES
Working range: 0.4100.0 IU/mL
PRECISION
Intra-lot Precision:Intra-lot precision was determined by testing of TSH reference materials using 10 test devices from the same lot. CV 15%.
Inter-lot Precision: Inter-lot precision was determined by testing of TSH reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV 20%.
ACCURACY
TSH control materials with two different concentrations were tested by every lot of test device, and the deviations were within 15%.
LINEARITY
A serial concentration of TSH reference materials at 0.4-100.0 IU/mL was tested, and the correlation coefficient (R) is 0.9900.