PRL FIA Test

Quantitative, non-subjective results.
Increased sensitivity and broader detection range compared to rapid tests.
RANGES
Working range: 5.0-4000.0 IU/mL
PRECISION
Intra-lot Precision
Intra-lot precision was determined by testing of PRL reference materials using 10 test devices from the same lot. CV 15%
Inter-lot Precision
Inter-lot precision was determined by testing of PRL reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV 20%.
ACCURACY
PRL control materials with three different concentrations were tested by every lot of test device, and the deviations were within 15%.
LINEARITY
A serial concentration of PRL reference materials at 5.0-4000.0 IU/mL was tested, and the correlation coefficient (R) is 0.9900.