Ferritin FIA Test
Quantitative, non-subjective results.
Increased sensitivity and broader detection range compared to rapid tests.
RANGES
Working range: 1.0-1000.0 ng/mL.
PRECISION
Intra-lot Precision.
Intra-lot precision was determined by testing of Ferritin reference. materials using 10 test devices from the same lot. CV 15%.
Inter-lot Precision.
Inter-lot precision was determined by testing of Ferritin reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV 20%.
ACCURACY
Ferritin control materials with three different concentrations were tested by every lot of test device, and the deviations were within 15%.
LINEARITY
A serial concentration of Ferritin reference materials at 2.0-1000.0 ng/mL was tested, and the correlation coefficient (R) is 0.9900.