MAU FIA Test

Quantitative, non-subjective results.
Increased sensitivity and broader detection range compared to rapid tests.
RANGES:
Working range: 2.0-300.0 mg/L.
PRECISION:
Intra-lot Precision:Intra-lot precision was determined by testing of MAU reference materials using 10 test devices from the same lot. CV 15%.
Inter-lot Precision:Inter-lot precision was determined by testing of MAU reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV 20%.
ACCURACY:
MAU control materials with three different concentrations were tested by every lot of test device, and the deviations were within 15%.
LINEARITY: