CK-MB FIA Test

Quantitative, non-subjective results
Increased sensitivity and broader detection range compared to rapid tests
RANGES
Working range: 1.0-80.0 ng/mL
PRECISION
Intra-lot Precision
Intra-lot precision was determined by testing of CK-MB reference materials using 10 test devices from the same lot. CV 15%.
Inter-lot Precision
Inter-lot precision was determined by testing of CK-MB reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV 15%.
ACCURACY
CK-MB control materials with three different concentrations were tested by every lot of test device, and the deviations were within 15%.
LINEARITY