Ferritin ELISA
Sensitivity:Detects ferritin levels as low as 0.06 ng/mL at 2SD.
Specificity:No cross reactivity detected with ASA at 10 mg/mL, bilirubin at 10 g/mL, FSH at 10,000 mIU/ML, hCG at 100,000 mIU/mL, hemoglobin at 0.1 mg/mL, lipids at 10 g/mL, LH at 1,000 mIU/mL, PRL at 5,000 ng/mL, respectively.
Accuracy:r= 0.997 against reference ELISA by an external lab.
Intra Assay Precision:CV 6.4%.
Inter Assay Precision:CV 9.3%.
Hook Effect:No hook effect up to 10,000 ng/mL.
Shelf life:18 months.
Open kit stability: 8 weeks at 2-8C, or until the expiration date.