PROG FIA Test
Quantitative, non-subjective results.
Increased sensitivity and broader detection range compared to rapid tests .
1. Range
Working range: 0.30-80.00 ng/mL
2. Precision
Intra-lot Precision
Intra-lot precision was determined by testing of PROG reference materials using 10 test devices from the same lot. CV 15%.
Inter-lot Precision.
Inter-lot precision was determined by testing of PROG reference. materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV 20%.
3. Accuracy
PROG control materials with two different concentrations were tested by every lot of test device, and the deviations were within 15%.
4. Linearity
A serial concentration of PROG reference materials at 3.00-80.00 ng/mL was tested, and the correlation coefficient (R) is 0.9900.
5. Specificity
Crossed Factors Concentration Test Result.
Estradiol 500 ng/mL < 0.3 ng/mL.
Testosterone 500 ng/mL < 0.3 ng/mL.
Cortisol 500 ng/mL < 0.3 ng/mL.
Aldosterone 500 ng/mL < 0.3 ng/mL.
Estriol 100 ng/mL < 0.3 ng/mL.
17- Estradiol 100 ng/mL < 0.3 ng/mL.
Oestrone 100 ng/mL < 0.3 ng/mL
17- hydroxyl progesterone 100 ng/mL < 0.3 ng/mL